Section 5: Software
EXAMPLE SET VALZO 1
>
DEFAULT 0
RANGE 0 or 1
COMMAND FCAL
TYPE Set/Get
DESCRIPTION Sets the number of points and the min/max pressure for eld calibraon. Points will be evenly
spaced between the minimum and maximum and will include a 0.00 point.
SYNTAX SET FCAL <number of points> <minimum pressure> <maximum pressure>
RETURNS Prompt when ready to accept a new command
EXAMPLE SET FCAL 13 -5.00 5.00
>
DEFAULT 0
RANGE These values are not range checked
COMMAND FVAL
TYPE Set/Get
DESCRIPTION Sets the number of points and the min/max pressure for eld validaon. Points will be evenly
spaced between the minimum and maximum and will include a 0.00 point.
SYNTAX SET FVAL <number of points> <minimum pressure> <maximum pressure>
RETURNS Prompt when ready to accept a new command
EXAMPLE SET FVAL 15 -5.00 5.00
>
DEFAULT 0
RANGE These values are not range checked