66
MPS4232
Section 5: Software
The oven sengs control the thermal chamber when performing a mul-temperature calibraon or validaon of the
MPS4200. These sengs are only required if performing a full calibraon or validaon. Refer to Secon 6: Calibraon and
Validaon Overview for more informaon.
DESCRIPTION Returns all of the oven sengs
SYNTAX LIST C
RETURNS All of the oven sengs followed by the prompt.
EXAMPLE LIST O
SET IPOVEN 0.0.0.0 0
SET STARTOVEN 0
SET STOPOVEN 0
SET TEMPOVEN 0
>
COMMAND IPOVEN
TYPE Set/Get
DESCRIPTION Sets the IP address and port for the oven / thermal chamber.
SYNTAX SET IPOVEN <ipadd> <port>
Where:
<ipadd> = Calibrator IP address
<port> = Calibrator port
If 0.0.0.0 0 is set, no oven is used.
RETURNS Prompt when ready to accept a new command.
EXAMPLE SET IPOVEN 10.0.0.122 1025
>
DEFAULT 0.0.0.0 0
RANGE These values are not range checked.
COMMAND STARTOVEN
TYPE Set/Get
DESCRIPTION Sets the command to send the oven / thermal chamber to start.
SYNTAX SET STARTOVEN [<start oven command>]
If no command is entered, a <CR> is sent to the oven.
RETURNS Prompt when ready to accept a new command.
EXAMPLE SET STARTOVEN RUNM
>
DEFAULT 0
RANGE These values are not range checked.