List of Setup Parameters VC 1100
Version 2 → parqalist/VC1100E
5-12 C01 / C02 - C11 / C12 © Schenck VIBRO GmbH, D-64273 Darmstadt, Juni 98
J 05 Channel A : Unit
Choises : g, m/s
2
, ips, mm/s, mils, µm
Select English or metric units for the measured vibration parame-
ter of channel A.
Select g or m/s
2
for vibration acceleration
Select ips or mm/s for vibration velocity
Select mils or µm for vibration dieplacement
Unit Measured Parameter
g m/s
2
mm/s ips
µm mils
J03 = a
J03 = v
J03 = s
If the unit does not match the selected measured parameter (J03),
the consistency check generates error-message -61.
J 06 Channel B : Unit
Choises : g, m/s
2
, ips, mm/s, mils, µm
Select English or metric units for the measured vibration parame-
ter of channel B.
Select g or m/s
2
for vibration acceleration
Select ips or mm/s for vibration velocity
Select mils or µm for vibration dieplacement
Unit Measured Parameter
g m/s
2
mm/s ips
µm mils
J04 = a
J04 = v
J04 = s
If the unit does not match the selected measured parameter (J04),
the consistency check generates the error-message - 61.