EasyManua.ls Logo

Shimadzu SPD-M20A - PDA Utility Service Menu; PDA Utility Starting Method; When Lcsolution Is Installed; When CLASS-VP Is Installed

Shimadzu SPD-M20A
108 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 6: Maintenance and Adjustments
SPD-M20A Service Manual
51
6.4 PDA Utility service menu
6.4.1 PDA Utility starting method
< When LCsolution is installed >
1) [Start], [Programs], [LabSolutions], [LCsolution], [LCMaintenance Tools] are chosen and
started.
LC Tool Launcher
2) PDA Utility is chosen with Application. The connected instrument name is chosen by
the Instrument No.
3) [Run] is clicked.
< When CLASS-VP is installed >
1) [Start], [Programs], [Chromatography], [PDA Utility] are chosen and started.
NOTE
When both of LCsolution are installed in one set of PC as CLASS-VP, PDA Utility
of CLASS-VP dose not start.
In this case, PDA Utility of CLASS-VP can be started by the following method,
using PDA Utility of LCsolution.
1) The [Start][Run] is chosen.
2) regedit is inputted into a name and [OK] is clicked.
3) The right click of the
“HKEY_LOCAL_MACHINE\SOFTWARE\Shimadzu\Labsolutions”
is chosen and carried out.
4) [Rename] is chosen and “Labsolutions” is changed into “Labsolutions_a” etc..
5) Please return after use to the original name.

Table of Contents

Related product manuals