Chapter 6: Maintenance and Adjustments
SPD-M20A Service Manual
67
Injector ............................................. Rheodyne 7125
Flow rate .......................................... 1 mL/ min.
Joint .................................................. For A and B, 1.6C couplings are used.
Between injector and Joint A ............ 0.3 2 m
jointless connection
NOTE
When performing injection, be careful not to allow bubbles to enter.
W is less than 10 seconds
(with INERT cell, less than 15
seconds)
b/a 2
Criteria
W: Peak width
b/a: Peak degree of warp
H
a
b
W
H
20