Functions
6.8 [2.09] Settings > Inputs/outputs
Operating with the Local User Interface
Operating Manual, 06/2017, A5E31930478-05
147
Table 6- 23 Function assignment at function (pre-selection): "Other functions"
Function
Trigger
Remark / effect
Switch pump
Positive/negative
edge
1)
Switch sample gas pump
3)
.
The pump is switched on at a positive edge and switched off at a negative
edge.
Measuring protection
Positive/negative
edge
1)
Activate measuring protection.
As long as the measuring protection is active, no other user actions on the
device are possible. This especially includes changes
to the configuration as
well as initialization of calibration via a digital input or via a bus interface.
A cyclical AutoCal that has already been configured is still executed, how-
ever because no user actions are necessary for this.
The measuring protection is only set if the following requirements are met:
• The device is in the "Measurement" operating mode.
• The inputs/outputs are not currently tested, for example, during the
commissioning/checking of communication with the control system:
→ Menu [3.11] Maintenance and diagnostics > Test
• The device is not in the "Function check" operating mode.
• No user is logged on.
The measuring protection is set with a delay under the following conditions:
• As long as function check is active
• As long as the operating mode is not "Measurement"
• As long as an already started calibration is still running (including expira-
tion of the sample gas purging time)
• If a calibration was triggered by a digital input, until the end of the cali-
bration or until the trigger is discarded.
• If an AutoCal/AutoVal is pending and its execution is delayed
• As long as an AutoCal/AutoVal is running, including sample gas inter-
mediate operation
Ack. curr. mess.
Positive edge
2)
Acknowledge current messages.
As long as the cause of a fault, a maintenance alarm, or a maintenance
demand has not been remedied, the corresponding message is still pend-
ing.
Messages are not acknowledged in the logbook but in the message list.
Software reset
Positive edge
2)
Software reset
1)
"Level-triggered": Activation by rising signal edge, deactivation by falling signal edge. Active at signal level "High".
2)
"Edge triggered": Activation by rising signal edge.
3)
Requirement: optional gas management module is installed.

Table of Contents

Other manuals for Siemens SIPROCESS GA700

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens SIPROCESS GA700 and is the answer not in the manual?

Siemens SIPROCESS GA700 Specifications

General IconGeneral
BrandSiemens
ModelSIPROCESS GA700
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals