Alarm, error, and system messages
7.3 ULTRAMAT 7 message list
Operating with the Local User Interface
268 Operating Manual, 06/2017, A5E31930478-05
Mes-
sage
Cause
Remedy
No.
Text
148 Saved user data faulty.
Content of user data in parameter
memory faulty and cannot be restored
Save the work data in the user data
again
Alternatively: transfer a valid parame-
ter set to the device.
149 Factory data faulty.
Content of factory data in parameter
memory faulty and cannot be restored
The saved factory data is faulty.
Contact Service.
150 Parameter value invalid.
Change of input limits of a parameter
due to change of another parameter.
The previously set value is now out-
side the changed input limits.
Due to a change in a parameter, the
value of another parameter is outside
its valid limits.
Correct the value of the associated
parameter.
151 Configuration restored
following fault.
Content of parameter memory was
faulty but has been recovered
An error was detected when loading
the device configuration.
However, a valid configuration could
be recovered without data loss.
153 Factory synchronization
of the electronics faulty.
Synchronization values of analyzer
module/option module electronics
faulty and cannot be restored
The
synchronization data saved at the
factory for the module electronics is
corrupt.
Replace the module.
238 Interference gas meas-
ured value uncertain or
faulty.
Check the sources for the measured
values of the parameterized interfer-
ence gases.
239 External pressure value
faulty.
Check the external pressure meas-
urement.