Alarm, error, and system messages
7.3 ULTRAMAT 7 message list
Operating with the Local User Interface
268 Operating Manual, 06/2017, A5E31930478-05
148 Saved user data faulty.
• Content of user data in parameter
memory faulty and cannot be restored
• Save the work data in the user data
again
• Alternatively: transfer a valid parame-
ter set to the device.
149 Factory data faulty.
• Content of factory data in parameter
memory faulty and cannot be restored
• The saved factory data is faulty.
• Contact Service.
150 Parameter value invalid.
• Change of input limits of a parameter
due to change of another parameter.
The previously set value is now out-
side the changed input limits.
• Due to a change in a parameter, the
value of another parameter is outside
its valid limits.
• Correct the value of the associated
parameter.
151 Configuration restored
following fault.
• Content of parameter memory was
faulty but has been recovered
• An error was detected when loading
the device configuration.
• However, a valid configuration could
be recovered without data loss.
153 Factory synchronization
of the electronics faulty.
• Synchronization values of analyzer
module/option module electronics
faulty and cannot be restored
• The
synchronization data saved at the
factory for the module electronics is
corrupt.
• Replace the module.
238 Interference gas meas-
ured value uncertain or
faulty.
• Check the sources for the measured
values of the parameterized interfer-
ence gases.
239 External pressure value
faulty.
• Check the external pressure meas-
urement.