EasyManua.ls Logo

Siemens SIPROCESS GA700 - Page 267

Siemens SIPROCESS GA700
306 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Alarm, error, and system messages
7.3 ULTRAMAT 7 message list
Operating with the Local User Interface
Operating Manual, 06/2017, A5E31930478-05
267
Mes-
sage
Cause
Remedy
No.
Text
75 Temperature sensor on
AM PCB faulty.
Electronic fault Replace the analyzer module printed
circuit board or the analyzer module.
80 Source current faulty.
Electronic fault Check the status of the source in
"[3.05.1.1] Measured signal".
If necessary, replace the analyzer
module.
81 Chopper faulty.
Electronic fault Replace the analyzer module.
82 Reference voltage outside
tolerance.
Electronic fault Replace the analyzer module.
83 Signal acquisition faulty
Electronic fault Replace the analyzer module.
84 Test signal outside toler-
ance.
Electronic fault Replace the analyzer module.
85 System error
Electronic fault Restart the device.
86 System error
Electronic fault Restart the device.
87 Parameter memory faulty.
Electronic fault Replace the module.
88 Analog output faulty.
Electronic fault Check the electrical circuit in the
menu "[3.11.1.1] Test analog out-
puts".
If necessary, replace the analyzer
module.
89 ADC faulty.
Electronic fault Replace the module.
95 Measured signal transient
reaction active.
Change the signal frequency
Turn the source off/on
Restart of chopper
Parameters of signal processing were
changed.
The measured value is invalid during
the subsequent settling time.
128 SPI communication faulty.
Electronics of the analyzer module or
the option module defective
Replace the module.
129 Communication with ADC
faulty
Analyzer module electronics defective Replace the analyzer module
144 System error
Internal software error Restart the device
145 Watchdog reset
If the error is signaled by the AM:
Analyzer module electronics defective
Internal software error (analyzer
module or processing module)
Acknowledge the message.
Contact Service.
146 Configuration faulty.
Content of parameter memory faulty
and cannot be restored
The active device configuration is
faulty.
Load the user or factory data in the
work data.

Table of Contents

Other manuals for Siemens SIPROCESS GA700

Related product manuals