Alarm, error, and system messages
7.3 ULTRAMAT 7 message list
Operating with the Local User Interface
Operating Manual, 06/2017, A5E31930478-05
267
Mes-
sage
Cause
Remedy
No.
Text
75 Temperature sensor on
AM PCB faulty.
• Electronic fault • Replace the analyzer module printed
circuit board or the analyzer module.
80 Source current faulty.
• Electronic fault • Check the status of the source in
"[3.05.1.1] Measured signal".
• If necessary, replace the analyzer
module.
81 Chopper faulty.
• Electronic fault • Replace the analyzer module.
82 Reference voltage outside
tolerance.
• Electronic fault • Replace the analyzer module.
83 Signal acquisition faulty
• Electronic fault • Replace the analyzer module.
84 Test signal outside toler-
ance.
• Electronic fault • Replace the analyzer module.
85 System error
• Electronic fault • Restart the device.
86 System error
• Electronic fault • Restart the device.
87 Parameter memory faulty.
• Electronic fault • Replace the module.
88 Analog output faulty.
• Electronic fault • Check the electrical circuit in the
menu "[3.11.1.1] Test analog out-
puts".
• If necessary, replace the analyzer
module.
89 ADC faulty.
• Electronic fault • Replace the module.
95 Measured signal transient
reaction active.
• Change the signal frequency
• Turn the source off/on
• Restart of chopper
• Parameters of signal processing were
changed.
• The measured value is invalid during
the subsequent settling time.
128 SPI communication faulty.
• Electronics of the analyzer module or
the option module defective
• Replace the module.
129 Communication with ADC
faulty
• Analyzer module electronics defective • Replace the analyzer module
144 System error
• Internal software error • Restart the device
145 Watchdog reset
• If the error is signaled by the AM:
Analyzer module electronics defective
• Internal software error (analyzer
module or processing module)
• Acknowledge the message.
• Contact Service.
146 Configuration faulty.
• Content of parameter memory faulty
and cannot be restored
• The active device configuration is
faulty.
• Load the user or factory data in the
work data.

Table of Contents

Other manuals for Siemens SIPROCESS GA700

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens SIPROCESS GA700 and is the answer not in the manual?

Siemens SIPROCESS GA700 Specifications

General IconGeneral
BrandSiemens
ModelSIPROCESS GA700
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals