3 — Installation Verification
19
10085896_H
V-PRO
®
maX 2
Operator Manual
Table 3-1. Cycle Limits (North America) (continued)
Cycle Name Weight Limit
lb (kg)
Device Type
Lumen ID
mm
Lumen
Length mm
Lumens Per
Load
Dead end lumens
≥ 1.3 ≤ 73
Table 3-2. Cycle Limits (Outside North America)
Cycle Name Weight Limit
lb (kg)
Device Type
Lumen ID
mm
Lumen
Length mm
Lumens Per
Load
Non Lumen
Cycle
50 (22.7)
Titanium or Stainless-Steel devices
with diffusion-restricted spaces
N/A N/A N/A
Non Lumen
Cycle
50 (22.7)
Non-lumened instruments including
non-lumened general medical
instruments, non-lumened rigid, semi-
rigid and flexible endoscopes
N/A N/A N/A
Fast Non
Lumen (if
equipped)
11 (5)
Titanium or Stainless-Steel devices
with diffusion-restricted spaces
N/A N/A N/A
Fast Non
Lumen (if
equipped)
11 (5)
Non-Lumened instruments and non-
lumened rigid, semi-rigid, and flexible
endoscopes.
N/A N/A N/A
Flexible Cycle
24 (11)
One single-channeled flexible
endoscope plus additional non-
lumened instruments
≥ 1.0 ≤ 1050
N/A
Two single-channeled flexible
endoscopes
One dual-channeled flexible endoscope
plus additional non-lumened
instruments
≥ 1.0
Lumen 1: ≤
990
Lumen 2: ≤
850
N/A
Two dual-channeled flexible
endoscopes
Lumen Cycle
19.65 (8.9)
Single-channeled stainless-steel
medical instruments
≥ 0.77 ≤ 500 20
Stainless-
Steel
Lumens
≥ 1.8 ≤ 542
Dual-channeled stainless-steel medical
instruments
≥ 0.77 ≤ 527
Dual-channeled stainless-steel medical
instruments
≥ 1.2 ≤ 275
≥ 1.8 ≤ 310
≥ 2.8 ≤ 317