12
2.3.1 Pull Report
The blue text box at bottom of the
display shows the progress of the
pull and then reports the final pull
results.
Figure 2-5. Pull Report.
2.3.2 View Pull Results
The Pull Results provides a means
by which one can review the
detailed results of a pull: Lines
used, number of loops, and Heat-
On-Time of the two most recent
pulls. To review or compare the pull
results, press [Menu] and then
[View Pull Results] to view the
report. Consistency of the Heat-On-
Time and comparing the behavior of
consecutive pulls can be useful tools
in developing stable parameter
settings.
Figure 2-6. View the pull report.
P-1000 FLAMING/BROWN MICROPIPETTE PULLER SYSTEM OPERATION MANUAL – REV. 3.02 (20161118)