EasyManua.ls Logo

Sutter Instrument P-1000 - Figure 4-5. Running the Ramp Test; Figure 4-6. Ramp Test Screen

Sutter Instrument P-1000
104 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
27
Press “Ramp” to run
a ramp test
Figure 4-5. Running the ramp test.
9. Now you will see the Ramp Test Screen.
Figure 4-6. Ramp Test screen.
The ramp test helps to establish safe heat settings. The ramp value is unknown for the
2.5 x 2.5 box filament (FB255B) filament in your puller and the 1.5 x 0.86 glass you will
be using, so you will need to run a Ramp test.
10. Install a piece of 1.5 x 0.86 capillary glass (BF150-86-10 sample glass was shipped with
your puller) and ADVANCE THE PULLER BARS ALL THE WAY TOGETHER before
tightening the knobs that clamp the glass. Do not over-tighten the knobs
.
CAUTION: The program HEAT value should not exceed the listed RAMP TEST
value by more than 10%.
The following shows how to load a piece of glass:
P-1000 FLAMING/BROWN MICROPIPETTE PULLER SYSTEM OPERATION MANUAL – REV. 3.02 (20161118)

Table of Contents

Related product manuals