P-1000 FLAMING/BROWN MICROPIPETTE PULLER SYSTEM OPERATION MANUAL – REV. 3.02 (20161118)
22
Lines with Heat = 0’. Please edit these lines.
Install a heat setting or erase this line of programming.
When Safe Heat is checked, A Ramp Value must be installed and the Heat setting must
be within 10% of the Ramp test Value.
Ramp = 0, Unable to Heat Check.
Press Ramp or uncheck Safe Heat (at your own risk).
Line 1 Heat > Ramp + 10%. Filament may burn out.
Reduce Heat or run Ramp Test, or uncheck Safe Heat (at your risk).
Line 1 Heat < Ramp - 10%. Jaws may overheat.
Increase Heat or run Ramp Test, or uncheck Safe Heat (at your risk).
Heat current error. Is Filament present & intact? See Menu/Diagnostics if Filament is
OK.
Filament Heater current is not behaving correctly. Make sure the filament is installed
and is not burned out or broken. Check the filament clamp screws to make sure they are
tight.
Failed to reach pressure
There may be an air leak or other problem. Go to Menu/Diagnostics to determine. Also,
see Home/Help Files/ Air system diagnostics to trace leaks.
Failed to melt glass.
This message shows if the glass does not soften after 60 seconds of heating. The Filament
may have burned out (go to Menu/Diagnostics) or the Heat value may be too low. Perform
a Ramp Test and use Safe Heat mode.