When the capillary glass is inserted,
even if the Pre-Heat temperature
has not been reached, the filament
heat is turned off to prevent the
glass from melting and/or sticking
to the filament.
If the jaw temperature is above
70°C when <PULL> is pressed, the
air flush occurs, but the pull cycle is
delayed until the jaws cool down to
70°C.
If the jaw temperature has dropped
to 70°C after the air flush, the pull
cycle will begin immediately.
P-1000 FLAMING/BROWN MICROPIPETTE PULLER SYSTEM OPERATION MANUAL – REV. 3.02 (20161118)