EasyManua.ls Logo

Tektronix TDS5000B Series - Limit Test for Performance Verification

Tektronix TDS5000B Series
237 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Commands
TDS5000B Series Service Manual
5-29
DIAg:RESults?
This query-only command returns an abbreviated status about the results of the
last diagnostic (or test) execution. For a more explanatory status message, use
the DIAg:RESults:VERBose query. This command is equivalent to selecting
Instrument Diagnostics from the Utilities menu, and then reviewing the
Diagnostic Status.
This query can be issued while diagnostics are sti l l in progress.
DIAg:RESults:VERBose?
DIAg:RESults?
DIAg:RESults?
This query might return
:DIAG:RESULTS “135”
indicating the specific failure code of the first test failure that occurred.
DIAG:RESults?
This query might return
:DIAG:RESULTS “Pass”
indicating the specific failure code of the first test failure that occurred.
DIAG:RESults?
This query might return
:DIAg:RESults “*****”
indicating the diagnostic test was not run.
DIAg:RESults:VERBose?
This query-only command returns a more explanatory message about the results
of the last diagnostic (or test) execution that the DIAg:RESults? query. This
command is equivalent to selecting Instrument Diagnostics from the Utilities
menu, and then reviewing the Diagnostic Status.
This query can be issued while diagnostics are sti l l in progress.