Thermo Scientific TSQ Altis, Quantis, and Fortis Hardware Manual 103
Maintaining the API Source Interface
Clean the ion sweep cone, the spray cone, and the ion transfer tube at the same time, since
they are closely connected.
Perform these tasks in order:
1. Removing the Ion Sweep Cone and the Ion Transfer Tube
2. Cleaning the Spray Cone and O-ring
3. Cleaning the Ion Transfer Tube
4. Cleaning the Ion Sweep Cone
Performing Source Maintenance Animation
1. Go to thermofisher.com.
2. In the search field, type TSQ Altis, TSQ Quantis, or TSQ Fortis.
3. Click the Catalog tab and then the instrument name.
4. Scroll down until you see the Product overview, Videos, and Documents tab.
Click the Videos tab and on the right, click Performing source maintenance.
CAUTION To prevent corrosion, do not use nitric acid to clean metal parts.
Tip You do not have to vent the system to remove the ion transfer tube.