■ Descriptions related to IN data
○ Gross weight (32 binary/0 to 99999)
○ Net weight (32 binary/0 to 99999)
Indicates gross weight or net weight. The weight status can be read at the same time, as in the example
below. The sign can be checked within the weight status. The weight displayed here is updated
constantly. No request signal etc. is needed. The order of HI and LO positions changes with
the
DWORD mode of F701+ (setting mode 6-1).
- ▲ Turns ON when at +1/4 scale division from the center point of each indicated scale
interval. Please note that it is necessary to set the 1/4 scale division (setting mode 3-7) to
ON.
(1) Zero calibration response
(2) Span calibration (actual load calibration) response
(3) Span calibration (equivalent calibration) response
11th word
15 14 13 12 11 10 9 8 7 6 5 4 3 2 1 0
(bit)
Command bit response
(3)(2) (1)
12th word
15 14 13 12 11 10 9 8 7 6 5 4 3 2 1 0
(bit)
Command no. response
(1) R/W response
(2) Command response
(3) Survival verification response (The survival bit status is reversed and set.)
(4) Remote READY ("1" is set after initialization.)
13th word
15 14 13 12 11 10 9 8 7 6 5 4 3 2 1 0
(bit)
Command bit response
(3)(2) (1)(4)
Nth word
15 14 13 12 11 10 9 8 7 6 5 4 3 2 1 0
(bit)
Weight
0 0
1) When DWORD mode is set to mode 0
0
RESERVE (always 0)
▼▲
CZ over
Sign Weight data (HI)
N+1th word
15 14 13 12 11 10 9 8 7 6 5 4 3 2 1 0
(bit)
Weight
Weight data (LO)
2) When DWORD mode is set to mode 1
Nth word
15 14 13 12 11 10 9 8 7 6 5 4 3 2 1 0
(bit)
Weight
Weight data (LO)
N+1th word
15 14 13 12 11 10 9 8 7 6 5 4 3 2 1 0
(bit)
Weight
0 0 0
RESERVE (always 0)
▼▲
CZ over
Sign Weight data (HI)