9 Specifications
141
Specifications
Chapter
9
9-4. Compliance with EC Directives
F701+ is a product which meets CE requirements. Follow the instructions below when using.
The F701+ Weighing Indicator is a CE-marked product that complies with EC Directives (based on the
European Community Council).
- Low voltage directive EN61010-1:2010
EN62311:2008 (test distance: 10cm)
- EMC Directive EN61326-1:2006
EN55011:2009, A1:2010 Group 1, Class A
EN61000-
3-2:2006, A1:2009, A2:2009
EN61000-3-3:2008
EN61000-4-2:2009
EN61000-4-3:2006, A1:2008, A2:2010
EN61000-4-4:2004, A1:2010
EN61000-4-5:2006
EN61000-4-6:2009
EN61000-4-8:2010
EN61000-4-11:2004
Pay attention to the following matters when installing the product.
1. Since F701+ is defined as an open type device (built-in device), be sure to install and fix it to a panel
and so forth before using.
2. The power cable
included in this product as standard may be used with an AC100V power supply in
Japan (nominal rated voltage AC125V).
When using this product overseas, use a power cable certified in that country.
3. Use shielded cables for cables other than the power cable (load cell, external I/O, options).
Overvoltage category II
Pollution level 2
          )
Caution
The combination of F701+ main unit and a lightning surge protector complies with
EN61000-4-5 (lightning surge immunity) of the EMC Directives.
EN62311 (human exposure) of the Low Voltage Directives is complied with at a
distance of 10cm.