27
3 Weighing
27
Weighing
Chapter
3
3-5-13.One-touch tare subtraction reset
This is a function that resets tare subtraction. If this operation is performed, tare weight can be cleared
using the one-touch tare subtraction operation.
1. Press the key.
2. Tare subtraction is reset.
3-5-14. Preset tare subtraction (Digital tare subtraction) (Setting mode 1-9, 2-3)
This is a function that subtracts a given setting
value from net weight
.
Preset tare subtraction (digital tare subtraction)
can be performed by setting tare value and
specifying the setting of preset tare weight to
"1: ON".
One-touch and preset tare subtraction operate
independently. Even when preset tare weight is
ON, one-touch tare subtraction is performed immediately after the key is pressed and the net
weight value becomes zero.
However, ON/OFF of preset tare weight as w
ell as setting of tare subtraction are prohibited when one-
touch tare subtraction is operating, depending on the limitation setting of tare subtraction function.
3-5-15.Limitation of the tare subtraction function (Setting mode 4-9)
Limitation can be imposed on the operation of
preset tare weight and one-touch tare
subtraction.
3-5-16.Tare weight display (Designated key)
This is a function that displays current tare weight. The tare weight that is displayed here is the result
including the one-touch tare subtraction and preset tare weight. This is displayed when "Tare weight
displayed with [0] key" of Measurement Act compliance (setting mode 4-9) is enabled.
- When tare weight is 0, " " goes out.
- When tare weight is other than 0, " " lights up.
- When (tare weight by one-touch tare subtraction) - (tare w
eight by preset tare weight) = 0, " "
goes out.
[Weighing function 3] (Setting mode 2-3)
1: Preset tare weight ON
0: Preset tare weight OFF
Preset tare weight
[Tare setting] (Setting mode 1-9)
(Input range: 0 to 99999)
[Measurement Act compliance] (Setting mode 4-9)
1: Enabled
0: Disabled
Tare weight displayed with [0] key
1: Enabled
0: Disabled
Prohibits the ON/OFF of tare setting
and preset tare weight when one-touch
tare subtraction is enabled.
Tare subtraction range
1: Tare Capacity
0: Total range
One-touch tare subtraction
1: Accept only when stable
0: Always accept