Smart Plus (PHT-35LHS)
Technical Manual
VT-TECH-Smart Plus-001
9 / 188
COPYRIGHT© 2018 VATECH.
ALL RIGHTS RESERVED
6. Image Optimization Command List for CT and PANO
Ensure to enter the following commands before starting an image optimization operation.
Smart Plus for CT and PANO
Modality Commands Description
CT
[SPM_CSON] CT Sensor Power On
[SPM_FISS_0001] Sensor Sensitivity & Binning mode setting
[SPM_FIPM_0001] Sensor Partial mode setting
// Sensor size: 0000~0032
[SPM_FISM_0001] Sensor Internal/External mode setting
//0000: Internal
//0001: External
[SPM_FOVO] Open all 4-axis collimators
[SPM_COLM_0009]
Collimator Position set
//0009: (AL) Close
Modality Commands Description
PANO
[SPM_PANO] Panoramic mode Initialization
[SPM_PSON] PANO sensor Power On
[SPM_FISS_0001] Sensor Sensitivity & Binning mode setting
[SPM_FIPM_0004]
Sensor Partial mode setting
// Sensor size: 0000~0032
[SPM_FISM_0001]
Sensor Internal/External mode setting
//0000: Internal
//0001: External
[SPM_FOVO] Open all 4-axis collimators
[SPM_COLM_0008]
Collimator Position set
//0008: (AL) Open