Smart Plus (PHT-35LHS)
Technical Manual
VT-TECH-Smart Plus-001
COPYRIGHT© 2018 VATECH.
6. CBCT CT Number Calibration
To optimize CBCT image quality in Smart Plus (PHT-35LHS) model, the operator should perform the
CBCT CT Number Calibration according to the following procedures.
Make sure that the operator should use TYPE A as CT PHANTOM JIG during the Smart
Plus (PHT-35LHS) CT Number Calibration.
Make sure that the operator should use TYPE A as CT NUMBER CHECK PHANTOM
during the Smart Plus (PHT-35LHS) CT Number Calibration.