Smart Plus (PHT-35LHS)
Technical Manual
VT-TECH-Smart Plus-001
78 / 188
COPYRIGHT© 2018 VATECH.
ALL RIGHTS RESERVED
9) Type the mA value that meets the standard of the Background Level in the High-Resolution
section. For Woman, subtract 0.3mA from the mA in Man mode and for Child, subtract 0.6mA
from the mA in Man mode and type the value in the section.
Do not change the kVps.
Make sure that each FOV (FOV 0, FOV 1, FOV 2) have the same values
By MAN_HARD=0950, 0870, ‘0950’ means 95kVp and ‘0870’ means 8.7mA.
10) Save and exit PwrParam_CBCT.ini.
ex) 8.7mA
ex) 8.4mA = 8.7mA - 0.3mA
ex) 8.1mA = 8.7mA - 0.6mA