Smart Plus (PHT-35LHS)
Technical Manual
VT-TECH-Smart Plus-001
COPYRIGHT© 2018 VATECH.
4. CBCT Rotator Position Alignment
This section covers the rotator's first position setup procedures by using the CT Number Phantom.
To optimize CBCT image quality in Smart Plus (PHT-35LHS), the operator should perform the
Chinrest Position Alignment operation according to the following standards and procedures:
Make sure that the operator should use TYPE A as CT PHANTOM JIG during the Smart
Plus (PHT-35LHS) CT Rotator Position Alignment.
Make sure that the operator should use TYPE A as CT NUMBER CHECK PHANTOM
during the Smart Plus (PHT-35LHS) CT Rotator Position Alignment.