1.4. Performing PANO Sensor Dark Calibration
1) Run VAKCAP.exe.
2) Click Settings. In the Serial port box, select a COM Port Number in use. Then click OK and
restart the program.
3) Click the Soft Calib tab and type [SPM_PANO], and then click Send to initiate the system into
PANO mode.