5. Background Calibration
To optimize CBCT image quality in Smart Plus (PHT-35LHS) model, the operator should perform the
CBCT Background Calibration according to the following standards and procedures.
5.1. CBCT Background Calibration Standard
CBCT Background Level Standard – Green mode
Exposure Initial-value – Green mode
- Man: 80 kVp, 5.0 mA
CBCT Background Level Standard – High-Resolution mode
Exposure Initial-value – High-Resolution mode
- Man: 95 kVp, 8.7 mA