Smart Plus (PHT-35LHS)
Technical Manual
VT-TECH-Smart Plus-001
169 / 188
COPYRIGHT© 2018 VATECH.
ALL RIGHTS RESERVED
0001 =>(SPM_CBCD_000x)CT_BEAM_CONTROL_DISABLE_SET
0: CT beam control enable in CT mode
1: CT beam control disable in CT mode
CT beam control setting in CT mode
0000: Control Enable In CT mode
0001: Control Disable in CT mode
000000000=>(spm_SNUM_xxxxxxxxx) SERIAL NUMBER
Product Serial Number
Ex) 000471234 means that the Product Serial Number is 047-1234
0001=>(spm_CFAS_000x)CONTROL_FILTER_APPLICATION_SET
0: Control filter is NOT applied
1: Control filter is applied
Control filter application setting value
0002 =>(SPM_MPTY_000x)MP3_TYPE_SELECT
0: D type(Original FW v1.00.16 ~ v1.00.18)
1: H type(Original FW ~v1.00.15)
2: J type(Original FW v1.00.19~)
MP3 file type select
0000=>(spm_GDAS_000x)GREEN_CT_DESCRIPTIVE_AUDIO_SET
0: Green CT descriptive audio OFF
1: Green CT descriptive audio ON
Green CT deactiptive audio on/off setting
0003=>(spm_MPOP_000x) MP3 MUSIC Option SET!!
0: Different music played each capture mode
1: Same music played each capture mode
2: Beep sound
3: Beep & AutoReturn sound (default)
Music setting During X-ray Exposure
0000: different music plays for each PANO/CT/CEPH mode
0001: same music plays for all PANO/CT/CEPH modes
0002: Beep sound plays
0003: Beep & AutoReturn sound plays (default)
0000=>(spm_MPEN_000x) MP3 Player board SET!!
MP3 Sound On/Off (Default value is 0001. But, if LCD is used for MP3 sound, the MPEN value would
be 0000.)
--- mp3 board enabled!!---
0075=>(spm_VOLU_00xx) MP3 Player board SET!!(1~82)
MP3 player volume setting value
~~~~~~~~~~~~~~~~~~~~~~~~~~~~
00000=>(spm_GCNT)PANO_CAPTURE_CNT
PANO Mode X-Ray Exposure Counting Value
00000=>(spm_GCNT)CEPH_CAPTURE_CNT
CEPH Mode X-Ray Exposure Counting Value
00000=>(spm_GCNT)ARCT_CAPTURE_CNT ct/ent
CT Mode X-Ray Exposure Counting Value
0001=>(spm_MRTP_x)MODE_RESET_TYPE
0: xyr axis reset when ceph mode change.
1: no xyr axis reset when ceph mode change.