Smart Plus (PHT-35LHS)
Technical Manual
VT-TECH-Smart Plus-001
83 / 188
COPYRIGHT© 2018 VATECH.
ALL RIGHTS RESERVED
6) Calculate the vertical center position (Slice XY(Z)) of CT Number Phantom between the bottom
position value and the top position value.
If Slice XY(Z) is 145(at TOP), 23(at BOTTOM), the median Slice XY(Z) value is 84.
Median Slice XY(Z) value = (145-23)/2 + 23= 61 + 23 = 84
7) Record the median Slice XY(Z) value.
8) Close the window and click V Slc.
TOP
BOTTOM