Cleaning
VTK-TR-BA-072e-1 - User Manual MDS 358X Series
Afterwards, use a fluid box cleaner. Check all bores
as well and then clean the rest of the fluid box
body.
Step 4 (cartridge base)
For the cartridge base, start with the fluidic brush
to clean all bores and openings of the cartridge
base.
Next clear the bore with a cleaning rod.
Finally use a fluid box cleaner to clean the wide
opening of the bore.
Important note!
If you use a cartridge base without integrated luer
lock, you need to clean the Fluid Box Connector
Luer Lock as follows. Otherwise, you can continue
with the next step.
Push a cleaning rod through the bore of the Fluid
Box Connector Luer Lock several times to remove
any residues of the medium.
Clean the center bore of the Fluid Box Connector
Luer Lock using a fluidic brush.
Clean the openings and thread of the Fluid Box
Connector Luer Lock using a fluid box cleaner. Use
it for the outside and the end of the bore.
Step 5 (tappet sealing)
For the tappet sealing, clean the bore of the tappet
sealing carefully with a fluidic brush.