14.6 Overview of the Command Functions
Details of the commands listed below are explained in paragraph 8.1.2.2, page 88.
RS-232C commands Functions
1
*ESR? (e.g. 50 errors) Check the latest error codes
2
*ESR2? (e.g. 50 errors) Check the latest error codes with parameters
3
*IDN? Check device information
4
*OPC? Check the last trigger impulses
5
ADJUST:? Check the adjust status
6
ADJUST:START Initiate the adjust
7
ADJUST:OFFSET:? Check current adjust offset value
8
ADJUST:OFFSET:<offset value> Set adjust offset value
9
ADJUST:CALIBRATION:? Check the status of (remote) adjust
10
ADJUST: CALIBRATION:ON Activate the adjust
11
HEATER:? Check the heater status
12
HEATER:ID? Check the information of the connected heater
13
HEATER:OFF Turn off the connected heater
14
HEATER:ON Turn on the connected heater
15
HEATER:TEMP:<target temperature> Set the target temperature for the connected heater
16
HEATER:LIMITS:<min>,<max> Set the limits for the temperature regulation
17
COOLER:? Check the cooling status
18
COOLER:ID? Check the information of the connected flow control valve
19
COOLER:OFF Turn off the connected flow control valve
20
COOLER:ON Turn on the connected flow control valve
21
COOLER:TEMP:<target temperature> Set the target temperature for the cooling
22
COOLER:OFFSET:<offset> Set the cooler offset for the connected flow control valve
23
KEY:ENTER Send ENTER signal to confirm error(s)
24
KEY:ESCAPE Send ESCAPE signal
25
HELP Check all RS-232C commands
26
LCD? Check current content of the screen
27
MAINT:STATUS Check accumulated amount of pulses (in %)
28
MAINT:MESSAGE:OFF Deactivate the maintenance message
29
MAINT:MESSAGE:ON Activate the maintenance message
30
SYSTEM:KLOCK:OFF Unlock the keypad
31
SYSTEM:KLOCK:ON Lock the keypad
32
SYSTEM:SHOW:CYCLES Check the current cycle counter
33
SYSTEM:SHOW:VALVEID Check the valve ID
34
SYSTEM:SHOW:CONTROLLERID Check the MDC ID
35
SYSTEM:SHOW:STATUS
Check the status of KeyLock, DosOK with Delay, SingleDosOK
and Auxiliary Mode
36
SYSTEM:SHOW:ACTTEMP Check the actuator temperature
37
SYSTEM:DOSOKDELAY:OFF Deactivate the DOSOK-Delay