Error Messages
VTK-TR-BA-072e-1 - User Manual MDS 358X Series
190 190 Incorrect Valve Data
This error message appears, if a checksum error occurs while writing the Cycle Counter.
Important Note!
The most common cause for this error is to unplug the valve before switching OFF the MDC or
before the MDC has completely shut down.
Error code display:
190 Incorrect Valve Data (Press Enter)
Error code status menu:
No entry
Error handling:
Press [enter]-key to acknowledge error message.
If error happens repeatedly, switch OFF control unit, and
inspect the connection. If necessary, exchange the cable
and/or valve.
Valve:
Valve remains unchanged. If the error message appears
during the start of the MDC, the valve is open; otherwise, it
is closed.
191 191 NozzleTappet Load Err.
During startup of the control unit, the system reads the counters (e.g. nozzle counter, tappet
counter). If a problem arises, this error message is displayed. Corresponding values may be
doubtful, until a new cycle is started.
Error code display:
191 NozzleTappet Load Err Enter.
Error code status menu:
No entry
Error handling:
Press [enter]-key to acknowledge error message. The
MDC restarts.
In case the error acknowledgement does not work, the
valve might be defect. If possible, test valve with
different MDC and MDC with different valve. If the valve
is defect, send it in for repair.
Valve:
Valve still open (start-up of the MDC)