Error Messages
VTK-TR-BA-072e-1 - User Manual MDS 358X Series
301 301 No Valve Present Error
The control unit does not recognize the valve during startup.
301 No Valve Present Error
301 No Valve Present Error
• The actuator connection has not been plugged in
correctly. Switch OFF MDC and inspect the actuator
cable and its connection. If the actuator cable is
damaged, it must be exchanged before restarting.
• Check MDC
• Send valve and/or MDC to VERMES Microdispensing or
to your supplier
Valve still open (start-up of the MDC)
302 302 Actuator Connection Error
With system in operation, the connection between valve (actuator) and MDC is disrupted.
302 Actuator Connection Error
302 Actuator Connection Error
• The actuator connection has not been plugged in
correctly. Switch OFF MDC and inspect the actuator
cable and its connection. If the actuator cable is
damaged, it must be exchanged before restarting.
• Check MDC
• Send valve to VERMES Microdispensing or to your
supplier
303 303 adjust error
This error can happen during operation. During the adjust a problem affects the calibration.
303 adjust error press enter
• Switch OFF MDC, switch it back ON and try again.
• If possible, check MDC with a different valve, to see, if
the adjust works there
• If possible, check valve with a different MDC, to see, if
the adjust works there
• Send valve and/or MDC to VERMES Microdispensing or
to your supplier