iVent201 Operator's Manual
7
CAUTION: Using the iVent
201
in combination with
devices such as humidifiers or filters can increase the
pressure gradient across the breathing system. Make sure
that such devices do not produce excessive resistance to
the airflow provided by the iVent
201
.
WARNING: Do not cover the iVent
201
while it is in use.
Make sure that the iVent
201
is positioned so that its inlet
ports and cooling vents are open to freely circulating air.
CAUTION: The iVent
201
must not be operated
immediately following storage or transport outside the
recommended operation condition
.
CAUTION: The iVent
201
internal battery contains lead
and must be disposed of in accordance with local
ordinances and environmental regulations.
CAUTION: Do not connect the O2 supply when
equipment is not in use.