iVent201 Operator's Manual
11
1.6 Specifications
1.6.1 Monitored Data Range, Resolution and
Accuracy
Parameter Range, resolution, accuracy
Breath type Range: Type: - Control, assist, spontaneous, sigh
Resolution: N/A
Accuracy: N/A
Respiratory
Rate
Range: 0 to 150/min
Resolution: 1/min
Accuracy: ±1 for 1 to 12/min
±2 for 12 to 150/min
Exhale tidal
volume
Range: 0 to 5000 mL
Resolution: 1 mL
Accuracy: ±10 mL or ±15% (whichever is greater)
for 0 to 300 mL
±10% for 300 to 5000 mL
Exhaled Minute
Volume
Range: 0 to 99.9L
Resolution: 0.1 L
Accuracy: ±10%
Peak flow Range: 0 to 140 L/min
Resolution: 1 L/min
Accuracy: ±10%