iVent201 Operator's Manual
223
Frequency Part Maintenance
1500 hours or 3
months of use
None O
2
Calibration
1500 hours or 3
months of use
Battery
Deep Discharge
and Recharge
Every 3 months of
storage
Internal Battery Recharge
Annually Air Inlet Filter
Cooling Air inlet
filter
Battery
Inlet muffler assy.
Outlet muffler assy.
Perform safety
checks
Replace listed parts
by an authorized
VersaMed Service
Technician
Every 15000 hours
Pneumatic
Assembly
Replace parts by
authorized
VersaMed Service
Technician
7.3 The O.V.T.
O.V.T. (Operational Verification Test) should be performed
each and every time a new patient circuit is used. The O.V.T.
procedure is covered in detail in section 4.1.2.