iVent201 Operator's Manual
141
Figure 4.34 The Adaptive Peak Flow pop-up
3. Turn the dial to select the
desired setting.
NOTE: The factory default value is “High.” VersaMed
recommends that you use “High.”
4. Press to accept the value. The
value will now be displayed
beside “Adaptive Peak Flow ”
in the Advanced Settings menu.
5. To return to the Main Window,
turn the control knob to select
“Close,” then press the knob.
4.3.6 Set Purge Interval
The iVent
201
periodically sends a burst of air through the sensor
tubes in order to clean them. The Advanced Settings menu
allows the user to determine the frequency of the automated
sensor purgings.
Note: This is an optional feature not available in all
units.
Possible values are 1, 2, 5 and 10 minutes, and Off. The factory
setting is 5 minutes.