iVent201 Operator's Manual
215
17. Auto Start alarm test. Restart
the unit with a test lung
connected to the patient circuit.
Wait for about one minute and
verify that the Auto Start alarm
is activated, and the unit starts
ventilation is Pressure Control
mode at the preset parameters.
Close the alarm pop up window
and switch the ventilator to
Standby.
6.4 The Sensor Failure Alarm
When the iVent
201
detects a Sensor Disconnect alarm or Check
Sensor condition it will enter Open Loop ventilation backup
mode. During Open Loop, ventilation is provided based on an
average of the previous inhaled tidal volumes (see Appendix
E).
If the Sensor Disconnect alarm triggers this pop-up
window: