EasyManua.ls Logo

VersaMed iVent201 - Guide to Adaptive Bi-Level

VersaMed iVent201
310 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Adaptive Bi-Level
180
5.2 Guide to Adaptive Bi-Level
Adaptive Bi Level is a support mode of ventilation,
where spontaneous breathing efforts are pressure
supported at a high (inspiratory) and low (expiratory)
pressure. In the absence of spontaneous breathing
efforts, the ventilator will provide mandatory breaths at
a rate indicated by the “respiratory rate” setting. The
default setting should be lower than the spontaneous
breathing rate to reduce dysynchrony with patient
breathing efforts.
Breaths can be terminated by a reduction in peak flow
to a pre-selected percentage. In high leak situations,
cycling to exhalation by reduced flow will be easier for
the patient if the %flow termination value is high. A
low percent peak flow termination value will mean
that the mandatory I-time will dominate the
inspiratory cycle. Therefore, if the % peak flow
parameter is set at a low level, the I-time setting must
be appropriate to avoid breath stacking.
Either Nasal CPAP masks or full facemasks may be
used with AB. Patients who breathe with their
mouths should use a full facemask or if not available,
a nasal mask with suitable chinstrap support. When
using higher inspiratory pressures it is also
recommended to use a full facemask to prevent the
escape of inflationary pressure via the oral cavity.
Please refer to Table 5.3 at the end of this section for
recommended masks.

Table of Contents

Other manuals for VersaMed iVent201