iVent201 Operator's Manual
39
CAUTION: When not using a Low Pressure O
2
Supply,
the Low Pressure O
2
Filter must NOT be used, as the
22mm port could easily be blocked.
When Low O
2
pressure is selected in the Advanced Settings
window, the analyzed FiO
2
level is displayed in the Main
window of the screen.
2.3 Breathing Circuit
The iVent
201
breathing circuit is disposable and intended only
for single-patient use. Never reuse or attempt to clean a single-
use breathing circuit or any of its components.
⇒ Only breathing circuit accessories approved by VersaMed
or authorized VersaMed dealers should be used with the
iVent
201
.
• The Inspiratory Limb consists of a connector
at one end of corrugated tubing, with a one-
way valve connected to the Patient Wye (also
referred to as the Wye Sensor) at the other
end.