iVent201 Operator's Manual
69
3.1.3 Standby and Patient Ventilation
When you have selected the weight, the iVent
201
will enter
Standby, ready to ventilate the patient with a press of the
control knob. By default iVent
201
starts up in SIMV Volume
control.
To start ventilation from Standby, return to the Main Screen,
rotate the control knob until the Start box is highlighted, then
press the control knob.
To return the ventilator to Standby mode: return to the Main
screen, turn the control knob to highlight the Mode Selection
box on the top right of the LCD display, and press the control
knob. At the bottom of the drop-down menu is Standby. Dial
the knob until Standby is selected, and press. The ventilator
will suspend operation, and can be started again by selecting
Start on the Main screen and pressing the control knob.
⇒ Always return the iVent
201
to Standby mode before
powering down.
3.1.4 Changing Ventilation Mode
The Mode Selection box is on the top right side of the LCD
display: