6.3 Alarms Tests
Alarms tests check the following alarms:
• Tube Disconnect
• Patient Disconnect
• Sensor Disconnect (Open Loop)
• AC Disconnect
• Low O2 Pressure
• Low Minute Volume
• High Pressure
• High O2%
• Low O2%
• Apnea
• High PEEP
• Auto Start
Warning: Do not perform these tests while the patient is
connected to the machine.
1. Connect the ventilator to AC
power and oxygen supply
according to the iVent
specification (refer to section
2.1.1 on page on page 28).
2. Power up the ventilator. It will
automatically run self-test.