EasyManua.ls Logo

VersaMed iVent201 - Alarms Tests

VersaMed iVent201
310 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
iVent201 Operator's Manual
209
6.3 Alarms Tests
Alarms tests check the following alarms:
Tube Disconnect
Patient Disconnect
Sensor Disconnect (Open Loop)
AC Disconnect
Low O2 Pressure
Low Minute Volume
High Pressure
High O2%
Low O2%
Apnea
High PEEP
Auto Start
Warning: Do not perform these tests while the patient is
connected to the machine.
1. Connect the ventilator to AC
power and oxygen supply
according to the iVent
specification (refer to section
2.1.1 on page on page 28).
2. Power up the ventilator. It will
automatically run self-test.

Table of Contents

Other manuals for VersaMed iVent201