Alarms
196
For as long as the condition causing
the alarm is not corrected, the
minimized alarm is colored red and
bears a text message describing the
alarm condition. If the alarm
condition is not resolved within two
minutes, the alarm pop-up will
maximize and sound an audible alert.
WARNING: Rectify all alarm conditions immediately.
Always make sure that the patient is safe. An alternate
means of ventilation must be available at all times.
If more than one alarm condition occurs, the alarm
messages will appear sequentially according to their
order of appearance from left to right on the minimized
alarm bar.
To clear a minimized green alarm indicator from the
bottom of the LCD screen, press clear for two seconds.
All alarms are logged in the logbook. Clearing an alarm
also creates a log entry in the logbook.
6.1.2 Alarm settings
The iVent
201
’s alarm settings are highly configurable. If you
need to change them, refer to Section 6.2, Alarm Settings.
Use ”Auto Settings” to adjust alarms to fit a particular
ventilation mode and parameter setting. See Section 4.2.1.
The following alarms are user-adjustable:
High and Low Respiratory Rate (to adjust this
rate, see Section 4.2.1.1)
High and Low Minute Volume (Section 4.2.1.2)