iVent201 Operator's Manual
197
• High and Low Pressure (Section 4.2.1.3)
The Low Pressure alarm is updated
automatically when setting the PEEP value. It
can also be adjusted from the Alarm Settings
window
• Apnea Time (Section 4.2.1.4)
• High and Low FiO
2
(Section 4.2.1.5)
This alarm is updated automatically when
setting the FiO
2
parameter and can also be
adjusted from the Alarm Settings window
• Leak % (Section 4.2.1.6)
The following alarm values are configurable:
• Volume Limit Reached
This alarm is automatically set from the
pressure controlled setting window, whereas
the other alarms are set from the Alarm
Settings window and can be adjusted from
the Tidal Volume window.
• Audio Alarm Volume
From 1-10: factory default is 8 (Section
4.2.3.1)
• Inverse I:E Ratio Alarm
On or Off: factory default is On (Section
4.2.3.2)
• Tidal Volume not delivered (Section 4.2.3.3)
Off or 15%-85% range: default is set to 85%
• Patient Circuit Disconnect (Section 4.2.3.4)