iVent201 Operator's Manual
117
6. When you have selected the
desired alarm limit, press the
control knob to confirm your
selection.
7. With the Leak slider highlighted
in relief, you may either press
the control knob to change the
Leak settings again, or dial the
control knob to select another
alarm setting to adjust, or select
Accept at the bottom right of the
Alarm Settings window to
accept all the alarm settings and
return to the Main Screen.
4.2.2 Auto Settings
The Auto Settings feature sets all the alarms in the Alarm
Settings window to suit the currently measured ventilation
parameters for the patient. Selecting Auto Settings will bracket
the measured values for each parameter in the Alarm Settings
window with appropriate limits for those parameters.
WARNING: The operator should carefully review and assess
the automatic alarm settings. Adjust alarms that are not
compatible with specific patient needs, or facility policy.
CAUTION: The factory default setting for “Leak Alarm” is
set to OFF. You must manually set the leak alarm to enable it.
To choose Auto Settings, bring up the Alarm Settings window
from the Main Menu. Turn the control knob until Auto
Settings is selected.