E.2.4 Parameters Setting
The following parameters shall be set:
Mandatory Breath Rate
O
2
%
Mandatory Tidal Volume
Peak Flow, Adaptive or set value
I Time, Adaptive or set value
Limit Pressure
Triggering type and sensitivity
Support pressure for pressure support breaths
E.3 Continuous Positive Airway Pressure
Mode
E.3.1 Definition
Continuous Positive Airway Pressure (CPAP ) is a ventilation mode intended for
patients who are breathing spontaneously at a rate sufficient to meet their
ventilation requirements. During CPAP, the airway pressure remains above
ambient at all times, reducing the work of breathing.
E.3.2 Available Breath Types
Patient spontaneous and pressure support breaths can be delivered in this mode.
Mandatory breaths are not given (except manual breaths).
E.3.3 Description