iVent201 Operator's Manual
221
7 Care, Maintenance, and Tests
This section shows you how to
• Clean and maintain the iVent
201
• Keep the iVent
201
operable with the help of
preventive maintenance
• Calibrate and perform diagnostic tests on the
iVent
201
7.1 Cleaning and Maintenance
While the iVent
201
has been designed to resist damage, staining,
and wear, taking care to perform an occasional cleaning and
following basic maintenance procedures will enhance its
lifecycle.
The following chart summarizes cleaning and routine
maintenance procedures and, where appropriate, time
intervals.
Table 7.1 Maintenance Procedures and Intervals
Part Procedure Comments
Ventilator
Wipe exterior clean with a
damp cloth and mild
detergent.
Do not allow
liquid to
penetrate into
the ventilator.
Cooling Vents &
Cooling Inlet
Filter
Clean every 1500 hours
(or 3 months) of use, or as
necessary.
Use vacuum
device to clean
the vents and
cooling air inlet
filter.