iVent201 Operator's Manual
13
Parameter Range, resolution, accuracy
RR/Vt-Rapid
Shallow
Breathing Index
(RSBI)
Range: 0 –200 /min*L
Resolution: 1 /min*L
Accuracy: ± (5 + 20% of actual value). /min*L
Static
compliance
Range: 0 to 99 ml/cmH
2
O
Resolution: 1 ml/cmH
2
O
Accuracy: ± (2 + 20 % of actual value). ml/cmH2O
Auto PEEP Range: 0 to 99 cmH
2
O
Resolution: 1 cmH
2
O
Accuracy: ± 2 (+ 5 % of reading) cmH
2
O
Time Constant Range: 0.0 to 9.9 sec
Resolution: 0.1 sec
Accuracy: ±20%
1.6.2 Size and Weight
Height: 13” / 33 cm
Width: 9.5” / 24 cm
Depth:
Screen:
10.3” / 26 cm
8.4” / 21.3 cm diagonal
Weight 15.4 lb / 7 kg (without battery)
Battery Weight 6.5 lb / 3 kg
Overall Weight 22 lb / 10 kg