iVent201 Operator's Manual
187
Rate
1 to 50 bpm depending on I time setting selected
FiO
2
21% to 100%
Rise Time
0.1 to 1.5 sec
Note: Rise time cannot exceed inspiration time.
P-High
2 to 60 cmH
2
O; with CAUTION indication at 41
cmH
2
O
Note: P-High cannot be set below P-Low +2.
P-Low
0 to P-High minus 2 or Maximum 30 cmH
2
O
Vpeak % Term
10% to 90% (default 40%)
I-Time
0.5 to 3 seconds; with CAUTION indication at
2.0 seconds or when rate setting I:E ratio
reaches 1:4
NOTE: If I-Time flashes, inspiration was
terminated by the set I-Time being reached.
When breath termination occurs due to
decreased flow, the I-Time display is steady.
Triggers:
Pressure
–0.5 to –20 cm (default –2)
Flow
1 to 20 L (default 2)
NOTE: Apnea Back Up ventilation operates in the same
manner as in other ventilation modes. See Appendix E4
for more details.
5.6 Adaptive Bi-Level Window
This is what the iVent
201
ventilating in Adaptive Bi-Level mode
looks like: