Alarms
208
Warning
Set Count or
Time of
Condition
Mode of Failure
Priority
Probable Cause Criterion for
Clearing Alarm
Corrective Action
Clinical situation
Volume
Limit
Reached
4 breaths
delay
Set volume limit
has been reached
in pressure control
modes
Mid Clinical situation
Inspiratory pressure
set too high
Volume limit is set
too low
Limit volume (V
T
)
is not reached
for 2 breaths
Assess patient
Reduce Inspiratory
pressure
Increase volume
limit
High PEEP
Alarm
2 breath delay
Actual PEEP value
more then 10 cm
from set value
High Exhalaton Valve
occlusion
Improper
parameter setting
PEEP pressure
decrease to set
value + 10
Check exhalation
valve
Check parameter
setting
Driver Failure A failure in the
Hold-Nebulizer
System
Low A short circuit in the
Hold-Nebulizer card.
N/A Remove unit for
inspection and
servicing by qualified
service personnel
O2 system
Failure
Abnormal O2 flow
or pressure.
High PSV Malfunctioning
N/A Remove the unit for
inspection and
servicing by qualified
service personal/