EasyManua.ls Logo

VersaMed iVent201 - Page 228

VersaMed iVent201
310 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Alarms
216
Figure 6.3 The Sensor Disconnect Alarm pop-up
Press the clear button.
Check the connection of flow transducer
tubing to the ventilator.
Then, check for excessive amounts of water or
secretions, which can set off the sensor alarm,
in the tubing and flow transducer.
If no water or secretions are visible, wait
approximately 10 breaths – sometimes
coughing can cause Check Sensor or Sensor
Disconnect alarms.
If the ventilator does not automatically return
to the previous mode and setup, press the
”Restore” option in the Backup Mode
Activated message and check the setting
values: PEEP, Rise-Time (auto).

Table of Contents

Other manuals for VersaMed iVent201