CAUTION: If the Check Sensor alarm persists
repeatedly, provide an alternative source of ventilation.
Then replace the patient circuit and run the O.V.T. test.
CAUTION: If the “Patient Disconnect” alarm is
available and switched off, the Sensor Failure alarm and
backup mode will not be activated.
NOTE: If the Sensor Disconnect alarm is detected three
(3) times within a minute, the ventilator will not restore
to the previous mode automatically, but will remain in
Open Loop instead. Press the “Restore” option in the
Backup Mode Activated pop-up to restore the ventilator
to the previous mode.