iVent201 Operator's Manual
205
Warning
Set Count or
Time of
Condition
Mode of Failure
Priority
Probable Cause Criterion for
Clearing Alarm
Corrective Action
Low SpO2 SpO
2
level is below
the set lower SpO
2
limit
High The patient clinical
condition, FiO
2
level
set too low, or SpO
2
alarm setting too
high
SpO
2
level is
higher than the
alarm limit.
Assessing the patient
clinical condition,
increasing the FiO
2
concentration, or
decreasing the SpO
2
alarm setting.
Auto Start Patient is
connected but the
unit is in Standby
mode
High
Mid
User enters Standby
mode unintentionally
User connects the
ventilator to the
patient but fails to
press Start
N/A Clear the alarm and
set the desired
ventilation
parameters
Service
Notice
Technical
irregularity (needs
Technical Support
Services)
High Power surge
Multiple sensor
disconnection
Calibration
required
Ventilator
Malfunction
N/A Turn the system
OFF and ON again
If the alarm is not
cleared, remove the
unit for inspection
and servicing by
qualified technical
personnel
Needs Cal
(Calibration)
System needs
calibration
Mid Calibrated values
irregularity
N/A Perform full
calibration by
qualified technician
Over
Temperature
Immediate
Internal
temperature of
ventilator has
exceeded 8C
High Cooling vents of the
ventilator are
blocked
High external heat
Malfunction of
internal cooling fan
Temperature
sensor failure
N/A Perform Preventive
Maintenance
procedures
Verify cooling vents
are clear and clean
Remove or protect
unit from external heat
source
Remove unit for
inspection and
servicing by